AccuPulse NIBP Simulator Technical Specifications

AccuPulse
AccuPulse NIBP CalTables™
Datascope Accutorr Plus
GE Medical Dash 4000
GE Medical Dinamap Pro
Philips Medical M3
Welch-Allyn Propaq Encore
Philips Medical VS1
Colin Medical
Criticare
CAS Medical
SunTech
Zoll Advatage
Smiths Medical Advisor
SpaceLabs SL 90207
SpaceLabs SL 90217
Welch Allyn VSM300
Welch Allyn VSM6000
Welch Allyn Spot VSM
Welch Allyn Spot LXi
Please contact the factory for the availability of other calibration tables.
BP Presets
Systolic/Diastolic (MAP):
ADULT:
250/190 (212)
200/150 (169)
150/100 (119)
120/80 (95)
100/70 (82)
80/50 (62)
60/30 (42)
NEONATAL:
150/120 (129)
120/90 (99)
100/70 (79)
80/50 (59)
60/30 (39)
35/15 (21)
Exact Diastolic, Systolic and MAP values are a function of the specific CalTables™.The BP Presets above are taken from the Welch-Allyn Propaq Encore CalTable™.
Accuracy:
<± 0.5% of selected target value
Oscillometric Pulse  
Pulse Amplitude @ MAP: 2.0 mmHg
Pulse Volume Range: 0 to 4.4 ml
Pulse Rise Time: 80 msec (minimum)
   
Pulse Rate  
Pulse Rate Range: 15-330 bpm
Pulse Rate Accuracy: ±1 bpm, 15-200 bpm,
±2 bpm 201-330 bpm
   
Displayed/MeasuredParameters
Dynamic NIBP Cuff PressureWaveform
Measurement Time: 0.0 to 180.0 seconds
Maximum Inflation Pressure: 0 to400 mmHg*
Inflation Time: 0.0 to 180.0 sec
   
Inflation Rate:
0 to 999.9 mmHg/sec
Minimum Pressure: 0 to 400 mmHg*
Deflation Time: 0.0 to 180.0 sec
Deflation Rate: -0.0 to -999.9 mmHg/sec
   
Digital Manometer
 
Pressure Range: 0.0 to 400.0 mmHg*
Accuracy: ± 0.5 mmHg, traceable to NIST
Resolution: ± 0.1 mmHg
   
Leak Test  
Start Pressure:
200 mmHg nominal,
10 to 370 mmHg via Remote Control
Pressure Drop: 0 to 400 mmHg*
Elapsed Time: 60 Seconds
High Resolution Manometer: 0.0 to 400.0mmHg*
   
Over-Pressure
Automatically tests operation of the monitor's relief valve
Pop-Off Pressure: 0 to 400 mmHg*
Instaneous Pressure: 0 to 400 mmHg*
   
Autosequences

User can utilize autosequences to test NIBP monitors with
a specific series of AccuPulse
NIBP performance tests.

 
Cuff Supports
Adult and Neonate mandrels are built into AccuPulse for portability and ease of use.
• Plastic and Aluminum Cylinders
• Adult (3.5" OD, 7.25" width)
• Neonate (1.25" OD, 2.25" width)
 
Adapter Hoses

Adapter hoses insert between the NIBP device,
cuff and analyzer. These adapters are compatible
with oscillometric NIBP monitors.
• Male/Female Luer
• Male/Female Clippard (GE Medical, Draeger/Siemens)
• Colder/CPC (GE Medical, Protocol Systems)
• OBAC Quick Release (Philips Medical)
• Universal 5/32" I.D. Hose

 
Self Test Accessories
• Pressure Bulb Assembly (tees into any Cuff Adapter)
• Self system leak test hose (plugged at distal end)
 
Communication Ports
2 RS-232 (second port available Q2 2006)
1 USB Port (available Q2 2006)
 
Display
Non-Glare Graphic LCD
CCFL Backlight
Resolution: 240x320 (1/4 VGA)
Thumbwheel contrast control
 
Power
100-240 VAC, 50 watts, 50-60 Hz, Desktop Switcher
Output: 24VDC @ 2.1A, 6 foot cable
Safety Agency Approvals: UL, CE,TUV
Internal Battery: nickel metal hydride (NiMH)
Voltage: 19.2V
Capacity: 2.5A-Hr, 150 NIBP Simulations
Charge Time: 16 Hours
 
Weight
8.0 pounds with internal battery
 
Dimensions
8" Wide x 5" High x 12" Long
 
Standard Accessories
• External Power Supply and Cord
• Adult and Neonate Cuff Supports (built-in)
• Operation Manual
• Internal Pump
• Internal Battery
 
Optional Accessories
• Carry Case
• Five Adapter Hoses
• Inflation Bulb with “T” fitting
• Self System Leak Test Hose

  *400.0 mmHg typical

 