SpO2 Simulator
Oxygen Saturation (SpO2)
| Range: |
0-100 SpO2% |
| Resolution: |
1 SpO2% |
| Accuracy: |
±0.5 SpO2%, 70-100 SpO2%
±1 SpO2%, 50-69 SpO2%
|
Pulse Rate
| Range: |
20-300 BPM |
| Resolution: |
1 BPM |
| Accuracy: |
±1 BPM, 20-199 BPM
±2 BPM, 200-300 BPM |
R-Curves (Red/IR Ratio)
R-Curves specify the relationship between the Red/IR ratio measured by pulse oximeters and the SpO2 value displayed . R-curves are obtained by each pulse oximeter manufacturer during clinical trials.
Preset Manufacturer's R-Curves (14): BCI, Criticare, CSI-Spot, Datascope, Datex, Hewlett-Packard (Philips), Masimo, Nellcor, Nellcor OxiMax, Nihon Kohden, Nonin, Novametrix, Ohmeda & Sensormedics
Please contact the factory for the availability of other R-Curves.
Pulse Modulation Amplitude
Simulates the effect of tissue perfusion
| Range: |
0.10 - 20.0% |
| Resolution: |
.05% |
Ambient Light
Simulates the ambient light level "seen" by the oximeter probe
| Range: |
DC Level & AC |
| Level: |
(5-255 DC) (0-255% AC)
|
| Resolution: |
1% |
| Accuracy: |
±1% |
| Flicker Frequency Presets (6): |
60 Hz |
|
120 Hz |
|
121 Hz |
|
50 Hz |
|
100 Hz |
|
101 Hz |
Motion Artifact
| Amplitude Range: |
0-100 |
| Frequency: |
4 Hz, Contact factory for availability of other frequencies. |
Arrhythmias
Simulates cardiac arrhythmias by altering the plethysmographic pulse amplitude and pulse rate.
| Arrhythmias (5): |
Asystolic Pause |
|
PVC |
|
Atrial Fibrillation |
|
Tachycardia |
|
Bradycardia |
Auto Sequences
Preset Auto Sequences automate pulse oximeter testing.
Standard Presets (18):
- All autosequences are user configurable
- User configurations are saved, replacing factory samples
Factory Samples
100-58 - preset for 13 manufacturers
50-81 - preset for Nellcor and Ohmeda
Trend Loop
4 Points
Pulse Rate: 40-200
Remote Control
The Smart Sat can be operated via remote control through its RS-232 port and duplicate all front panel operations.
Probe Analyzer
Probe Continuity Test
Simultaneous Channels:
2 for Nellcor-style probes: LED & PhotoDiode circuits
3 for Ohmeda-style probes: Red LED, IR LED & PhotoDiode circuits
Test Mode: Constant Current Stimulus, Voltage Output
Constant Current Stimulus: 500 uA
Intermittent "Glitch" Detection: Detects open circuits and short circuits, 1 msec or longer
Probe Connection: Connectors for Nellcor & Ohmeda probes are built into SmartSat; other probes are connected via adapter cables
General
Display
Type: Backlit graphic LCD
Resolution: 320 h x 240 v, 0.30mm dot pitch
Controls: Contrast
Serial Interface
Remote Control Mode
Type: RS-232C, 9 pin "PC-AT" style
Baud Rates: 1200 bps
Battery & AC Adapter
Battery
Type: Nickel Metal Hydride (NiMH)
Voltage: 19.2V
Capacity: 2.1A-Hr (About 4.5 Hours usage)
Charge Time: 17 Hours (charges even if SmartSat in use)
24 VDC External AC Adapter : 120 VAC / 60 Hz, 38 watts
or, 230 VAC / 50 Hz, or 100 VAC adapter as appropriate
Physical
Case: Aluminum with non-skid feet & handle
Dimensions: 8" Wide x 8" Deep x 5" High
Weight: 4 pounds including battery
SmartSat Cables
The following Simulator and Probe Analyzer cables are available for purchase.
Please contact Clinical Dynamics for details.
SpO2 Simulator Adapter Cables: BCI, Criticare, CSI-Spot, Datascope, Datex, Datex-Ohmeda, Philips (HP), Masimo, Nellcor / OxiMax, Nihon Kohden, Nonin, Novametrix, Sensormedics, & Tuffsat
Probe Analyzer Adapter Cables: Criticare, Datascope, Datex, Philips, Masimo, Nihon Kohden, Novametrix & Sensormedics
Note: BCI, Nellcor, Nonin & Ohmeda probes connect directly.
Optional Accessories
SmartSat Carry Case
|